Machinery Balancing Module
How to Perform a Single-Plane Procedure
Add a Trial Weight
• From the Initial Run screen, press an Enter button. The Balancer – Trial Weight
TW A1 screen displays.
Figure 5 - 8.
The Trial Weight Screen.
• Stop the machine.
• On the Microlog, enter the Trial Weight, Angle and Radius for the trial weight you
are attaching to the machine.
The Rotor Mass field is used if you wish to have the Microlog
estimate the trial weight. Reference the How to Estimate Trial
Weights section below for detail.
Trial weights are typically attached at zero degrees (the trigger
reference).
Unless you wish the Microlog to estimate a trial weight, the
Radius entry is not necessary, so long as you position the trial,
correction, and trim weights at the same radius.
You can enter a negative number if the trial weight has been
removed instead of added.
• On the machine, carefully attach the precise weight at the precise angle and radius
specified. Balancing results depend greatly on the precision of your measurements
and actions.
5 - 14 SKF Microlog - GX Series
User Manual