Introduction to the SKF GX Series Microlog
Battery Operations
CriticalVery limited battery life remaining (less than 45-60 minutes).
Charge the battery immediately, it is not recommended to continue work when
the battery status is Critical.
ChargingIndicates the battery is charging.
IMPORTANT: Connect the Microlog to its Power Supply/Battery Charger
whenever you do not intend to resume operations in the immediate future.
Database damage may occur if battery power is completely discharged while the
unit is in operation or in suspend mode.
How to Install the Battery Pack
IMPORTANT: To ensure proper and safe Microlog operation, ONLY use the
supplied battery pack.
To install the Lithium-Ion battery pack in your Microlog:
Remove the two screws on the battery cover located on the underside of your
device and lift the battery cover away from the device.
The main battery is removed by sliding the cell to the left and lifting it out of the
device.
Insert the replacement lithium-ion battery in the battery compartment and slide to
the right to fit into place.
Replace the battery cover and press down firmly on the cover while re-securing the
latches.
WARNING! To minimize the risk of serious injury or damage:
Do not expose the battery pack to temperatures in excess of 140 °F (60 °C).
Do not disassemble, incinerate, or short-circuit the battery pack.
Danger of explosion if battery is incorrectly replaced.
Replace the battery only with the same or equivalent type recommended by the
manufacturer. Dispose of used batteries according to the manufacturer’s
instructions.
Do not open battery compartment in a hazardous / explosive area.
Replacing the Battery in an Active Microlog
If the Microlog is running from battery power only, removing the battery causes the
Microlog to shut off power to the LCD display and to the SD card. The Microlog may
lose any data stored in RAM. This could result in database errors!
To avoid these problems, use the following procedure to replace the battery pack in an
active Microlog running on battery power:
Press the Power button for two seconds to suspend the Microlog.
Unlatch the battery and remove the pack.
1 - 20 SKF Microlog - GX Series
User Manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SKF Microlog GX Series and is the answer not in the manual?

SKF Microlog GX Series Specifications

General IconGeneral
BrandSKF
ModelMicrolog GX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals