Spindle Test Module
Spindle Tests Overview
Place the indicator in the bottom and outer position of the taper, approximately 1/4
from the ends.
Figure 1015.
Indicator Position.
Slowly rotate the shaft manually and read the total range value. Key in the result.
Figure 1016.
Enter Run-Out Range Value.
HSK tapers: after completing the radial measuring, follow up with an axial measure
of contact face and evaluate against the acceptance standard.