Frequency Response Function Module
Saving FRF Data
Field Calibration Set-up Fields
Access the FRF – Setup screen following the instructions under the heading Set-up
the FRF Module, earlier in this chapter.
Set Measurement to Apparent Mass.
Set Meas Units to N/m/s
2
.
The Microlog should then display 1 kg (or other chosen mass),
plus the weight of the accelerometer:
M = calibrated mass + accelerometer mass (e.g., 1 kg + 10.5 g =
1.0105 kg).
Saving FRF Data
Saving, Reviewing, and Deleting Measurements
Procedures for saving measurement setups, for saving measurement results, for
viewing stored measurements, and for deleting stored measurements are similar
between modules. Reference Chapter 4, The Analyzer Module section on How to
Save, View, and Delete Analyzer Measurement Results for detailed information on
these procedures.