Non ROUTE
How to Collect Pre-set Non ROUTE Measurements
For Accel, Accel Time, Velocity, and gE pre-set measurement types, the following
standard settings are taken from the current selected ROUTE measurement:
Sensitivity
Coupling
Detection
Window
Accel Performs a “standard” type vibration Acceleration measurement.
Measurement parameters are set to:
Units - Accel (Gs)
Display FormatSpectrum
High-pass frequency - 1.1 Hz
Fmax - 10 kHz
Lines/Samples - 800 Lines
Averages – 4
Input Channel CH1
Overlap50%
Ext. Trigger Slope Positive
Trigger Level 2 V
Bin Zeroing On
Type - Spectral
Accel Time Performs the same “standard” type vibration Acceleration measurement,
but collects time waveform data. Measurement parameters are set to:
Units - Accel (Gs)
Display FormatTime
High-pass frequency - 1.1 Hz
Fmax - 8 kHz
Lines/Samples6400 Lines / 16384 Samples
Averages4 (time sync.) If a trigger is defined for the active ROUTE point, if
no trigger is detected only 1 average is taken.
Input ChannelCH1
Overlap50%
Ext. Trigger Slope Positive
Trigger Level 2 V
Bin ZeroingOn
Type Time
VelocityPerforms astandard” type acceleration to velocity vibration measurement.
Measurement parameters are set to:
UnitsA -> V
Display FormatSpectrum
High-pass frequency - 1.1 Hz
Fmax - 1 kHz
Lines/Samples - 400 Lines
Averages – 4
Input Channel CH1
Overlap50%
A - 4 SKF Microlog - GX Series