4. Troubleshooting
40900
40900
Attention message: F replenishment
No. 00909[F]
Replenishment tank is empty. Cannot start the operation.
Attention message release
YES
Condition
Check Point
Diagnosis
NOTE
• To access the wiring diagrams, click Wiring diagrams in Bookmarks.
Condition
When trying to measure the output amount or perform the initial replenishment when the replenishment solution level sensor for
replenishment tank is under the conditions below.
Suffix number Condition
00001 P1R (Lower) Replenishment Solution Level Sensor: OFF
00002 P2RA (Lower) Replenishment Solution Level Sensor: OFF
00004 P2RB (Lower) Replenishment Solution Level Sensor: OFF
00008 PSR (Upper) Replenishment Solution Level Sensor: OFF
Create the replenishment solution. -
Check each replenishment solution level sensors status via Input Check.
☞35220
Failed part Manual No.
• P1R (Lower) Replenishment Solution Level Sensor
• P2RA (Lower) Replenishment Solution Level Sensor
• P2RB (Lower) Replenishment Solution Level Sensor
• PSR (Upper) Replenishment Solution Level Sensor
☞63280
F replenishment I/O PCB
☞66800
Processor control PCB ☞66200