Error message: F replenishment
1/16
4. Troubleshooting
45900
45900
Error me ssage: F replen ishment
No. 05900[F]
Replenisher cartridge open motor operation error.
Error message release
NO
IMPORTANT
• If the above error message is shown, do not click YES to close it and continue the printing process without performing the
diagnosis below.
After a set number of prints are output,
☞No. 05912[F] Predetermined process amount was exceeded. will be shown,
and then printing cannot be performed.
NOTE
• If printing process is continued after the above error is shown, the time is counted as Replenishment Lack Time and the P1R, P2RA
and P2RB replenishers will not be replenished.
Condition
Suffix number Condition
00001 When the replenisher cartridge setting part is moved to the lower position for opening the replenisher cartridge, the
Replenisher Cartridge Position Sensor (Upper) does not turn to LIGHT even after a specified time has passed.
00002 When the replenisher cartridge setting part is moved to the lower position for opening the replenishment cartridge, the
Replenisher Cartridge Position Sensor (Lower) does not turn to DARK even after a specified time has passed.
00003 When the replenisher cartridge setting part is moved to lower position to open the replenishment cartridge, Replenisher
Cartridge Position Sensor (Lower) detects the error.
Although the replenisher cartridge setting part has been moved upwards then, Replenisher Cartridge Position Sensor
(Lower) does not turn to LIGHT even after a specified time has passed.
(When restarting after operation stop such as instant-off)
00004 When the replenisher cartridge setting part is moved to the upper position after the replenishment cartridge was failed
to be opened, the Replenisher Cartridge Position Sensor (Lower) does not turn to LIGHT even after a specified time
has passed.
00005 When the replenisher cartridge setting part is moved to the upper position after the replenishment cartridge was failed
to be opened, the Replenisher Cartridge Position Sensor (Upper) does not turn to DARK even after a specified time has
passed.
00006 When the replenisher cartridge setting part is moved to upper position due to the failure in opening the replenishment
cartridge, Replenisher Cartridge Position Sensor (Upper) detects the error.
Although the replenisher cartridge setting part has been moved downwards then, Replenisher Cartridge Position Sensor
(Upper) does not turn to LIGHT even after a specified time has passed.
(When restarting after operation stop such as instant-off)
00007 When the replenisher cartridge setting part is moved to the cap cleaning position for cleaning the cap, the Replenisher
Cartridge Position Sensor (Lower) does not turn to LIGHT even after a specified time has passed.
00008 When the replenisher cartridge setting part is moved to the cap cleaning position for cleaning the cap, the Replenisher
Cartridge Position Sensor (Upper) does not turn to DARK even after a specified time has passed.
00009 When the replenisher cartridge setting part is moved to the cap cleaning position to clean the cap, Replenisher Cartridge
Position Sensor (Upper) detects the error.
Although the replenisher cartridge setting part has been moved downwards then, Replenisher Cartridge Position Sensor
(Upper) does not turn to LIGHT even after a specified time has passed.
(When restarting after operation stop such as instant-off)
00010 When the replenisher cartridge setting part is moved to the lower position for draining the water after cleaning the cap,
the Replenisher Cartridge Position Sensor (Upper) does not turn to LIGHT even after a specified time has passed.
00011 When the replenisher cartridge setting part is moved to the lower position for draining the water after cleaning the cap,
the Replenisher Cartridge Position Sensor (Lower) does not turn to DARK even after a specified time has passed.
00012 When the replenisher cartridge setting part is moved to the lower position for draining the water after cleaning the cap,
Replenisher Cartridge Position Sensor (Lower) detects the error.
Although the replenisher cartridge setting part has been moved upwards then, Replenisher Cartridge Position Sensor
(Lower) does not turn to LIGHT even after a specified time has passed.
(When restarting after operation stop such as instant-off)
00013 When the replenisher cartridge setting part is moved to the upper position after the replenishment cartridge is opened or
cap is cleaned, the Replenisher Cartridge Position Sensor (Lower) does not turn to LIGHT even after a specified time
has passed.