:TRIGger Commands 35
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1325
:TRIGger[:EDGE]:COUPling
(see page 1610)
Command Syntax
:TRIGger[:EDGE]:COUPling <coupling>
<coupling> ::= {AC | DC | LFReject}
The :TRIGger[:EDGE]:COUPling command sets the input coupling for the selected
trigger sources. The coupling can be set to AC, DC, or LFReject.
• AC coupling places a high-pass filter (10 Hz for analog channels, and 3.5 Hz for
all External trigger inputs) in the trigger path, removing dc offset voltage from
the trigger waveform. Use AC coupling to get a stable edge trigger when your
waveform has a large dc offset.
• LFReject coupling places a 50 KHz high-pass filter in the trigger path.
• DC coupling allows dc and ac signals into the trigger path.
Query Syntax
:TRIGger[:EDGE]:COUPling?
The :TRIGger[:EDGE]:COUPling? query returns the current coupling selection.
Return Format
<coupling><NL>
<coupling> ::= {AC | DC | LFR}
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":TRIGger:MODE" on page 1308
• ":TRIGger[:EDGE]:REJect" on page 1327
The :TRIGger[:EDGE]:COUPling and the :TRIGger[:EDGE]:REJect selections are coupled.
Changing the setting of the :TRIGger[:EDGE]:REJect can change the COUPling setting.