312 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
10 :CHANnel<n> Commands
:CHANnel<n>:PROBe:HEAD[:TYPE]
(see page 1428)
Command Syntax
:CHANnel<n>:PROBe:HEAD[:TYPE] <head_param>
<head_param> ::= {SEND0 | SEND6 | SEND12 | SEND20 | DIFF0 | DIFF6
| DIFF12 | DIFF20 | NONE}
<n>::={1|2|3|4}
The :CHANnel<n>:PROBe:HEAD[:TYPE] command sets an analog channel probe
head type and dB value. You can choose from:
SEND0 — Single-ended, 0dB.
SEND6 — Single-ended, 6dB.
SEND12 — Single-ended, 12dB.
SEND20 — Single-ended, 20dB.
DIFF0 — Differential, 0dB.
DIFF6 — Differential, 6dB.
DIFF12 — Differential, 12dB.
DIFF20 — Differential, 20dB.
Query Syntax
:CHANnel<n>:PROBe:HEAD[:TYPE]?
The :CHANnel<n>:PROBe:HEAD[:TYPE]? query returns the current probe head
type setting for the selected channel.
Return Format
<head_param><NL>
<head_param> ::= {SEND0 | SEND6 | SEND12 | SEND20 | DIFF0 | DIFF6
| DIFF12 | DIFF20 | NONE}
See Also "Introduction to :CHANnel<n> Commands" on page 303
":CHANnel<n>:PROBe" on page 311
":CHANnel<n>:PROBe:ID" on page 313
":CHANnel<n>:PROBe:SKEW" on page 316
":CHANnel<n>:PROBe:STYPe" on page 317
NOTE