:TRIGger Commands 33
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1187
:TRIGger:NFC:TIMeout:ENABle
(see page 1428)
Command Syntax
:TRIGger:NFC:TIMeout:ENABle {1 | ON}
The :TRIGger:NFC:TIMeout:ENABle command enables the timeout period.
Currently, ON is the only valid setting.
With the ATRigger (Arm & Trigger) trigger event, if the second event does not
occur within the timeout period, the oscilloscope triggers when the timeout period
expires.
Query Syntax
:TRIGger:NFC:TIMeout:ENABle?
The :TRIGger:NFC:TIMeout:ENABle? query returns the timeout period enable
setting.
Return Format
<setting><NL>
{1}
See Also ":TRIGger:NFC:AEVent" on page 1180
":TRIGger:NFC:ATTime" on page 1181
":TRIGger:NFC:SOURce" on page 1182
":TRIGger:NFC:STANdard" on page 1183
":TRIGger:NFC:TEVent" on page 1184
":TRIGger:NFC:TIMeout" on page 1186
":TRIGger:NFC:TIMeout:TIME" on page 1188