:SBUS<n> Commands 29
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 773
:SBUS<n>:A429:TRIGger:PATTern:SDI
(see page 1428)
Command Syntax
:SBUS<n>:A429:TRIGger:PATTern:SDI <string>
<string> ::= "nn" where n ::= {0 | 1 | X}, length always 2 bits
The :SBUS<n>:A429:TRIGger:PATTern:SDI command defines the ARINC 429
two-bit SDI pattern resource according to the string parameter. This pattern
controls the SDI pattern searched for in each ARINC 429 word.
The specified SDI is only used if the :SBUS<n>:A429:FORMat includes the SDI
field.
Query Syntax
:SBUS<n>:A429:TRIGger:PATTern:SDI?
The :SBUS<n>:A429:TRIGger:PATTern:SDI? query returns the current settings of
the specified ARINC 429 two-bit SDI pattern resource in the binary string format.
Return Format
<string><NL> in nondecimal format
Errors "-241, Hardware missing" on page 1387
See Also "Introduction to :TRIGger Commands" on page 1139
":SBUS<n>:A429:FORMat" on page 767
":SBUS<n>:A429:TRIGger:TYPE" on page 776
":SBUS<n>:A429:TRIGger:PATTern:DATA" on page 772