EasyManua.ls Logo

Keysight Technologies InfiniiVision MSO-X 4154A

Keysight Technologies InfiniiVision MSO-X 4154A
1580 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
710 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
26 :POWer Commands
:POWer:TRANsient:IINitial
(see page 1428)
Command Syntax
:POWer:TRANsient:IINitial <value>[suffix]
<value> ::= Initial current value in NR3 format
[suffix] ::= {A | mA}
The :POWer:TRANsient:IINitial command to specify the initial load current value.
The initial load current will be used as a reference and to trigger the oscilloscope.
Query Syntax
:POWer:TRANsient:IINitial?
The :POWer:TRANsient:IINitial query returns the initial load current value.
Return Format
<value><NL>
<value> ::= Initial current value in NR3 format
See Also ":POWer:SIGNals:VSTeady:TRANsient" on page 697
":POWer:TRANsient:APPLy" on page 708
":POWer:TRANsient:EXIT" on page 709
":POWer:TRANsient:INEW" on page 711
":POWer:TRANsient:NEXT" on page 712

Table of Contents