:MARKer Commands 21
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 463
:MARKer:XUNits
(see page 1428)
Command Syntax
:MARKer:XUNits <units>
<units> ::= {SEConds | HERTz | DEGRees | PERCent}
The :MARKer:XUNits command sets the X cursors units:
SEConds — for making time measurements.
HERTz — for making frequency measurements.
DEGRees — for making phase measurements. Use the :MARKer:XUNits:USE
command to set the current X1 location as 0 degrees and the current X2
location as 360 degrees.
PERCent — for making ratio measurements. Use the :MARKer:XUNits:USE
command to set the current X1 location as 0 percent and the current X2
location as 100 percent.
Changing X units affects the input and output values of the :MARKer:X1Position,
:MARKer:X2Position, and :MARKer:XDELta commands/queries.
Query Syntax
:MARKer:XUNits?
The :MARKer:XUNits? query returns the current X cursors units.
Return Format
<units><NL>
<units> ::= {SEC | HERT | DEGR | PERC}
See Also "Introduction to :MARKer Commands" on page 453
":MARKer:XUNits:USE" on page 464
":MARKer:X1Y1source" on page 458
":MARKer:X2Y2source" on page 461
":MEASure:SOURce" on page 529
":MARKer:X1Position" on page 457
":MARKer:X2Position" on page 460