816 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:CXPI:TRIGger:PTYPe
(see page 1428)
Command Syntax
:SBUS<n>:CXPI:TRIGger:PTYPe {{0 | OFF} | {1 | ON}}
For the trigger types that let you trigger on data, the
:SBUS<n>:CXPI:TRIGger:PTYPe command specifies whether you want to trigger
when the special PTYPE byte is present (ON) or not present (OFF).
Query Syntax
:SBUS<n>:CXPI:TRIGger:PTYPe?
The :SBUS<n>:CXPI:TRIGger:PTYPe? query returns the PTYPE trigger setting.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also • ":SBUS<n>:CXPI:TRIGger" on page 813
• ":SBUS<n>:CXPI:TRIGger:IDFilter" on page 815
• ":SBUS<n>:CXPI:TRIGger:PATTern:DATA" on page 817
• ":SBUS<n>:CXPI:TRIGger:PATTern:DATA:LENGth" on page 818
• ":SBUS<n>:CXPI:TRIGger:PATTern:DATA:STARt" on page 819
• ":SBUS<n>:CXPI:TRIGger:PATTern:ID" on page 820
• ":SBUS<n>:CXPI:TRIGger:PATTern:INFO:CT" on page 821
• ":SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC" on page 822