:SBUS<n> Commands 29
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 847
:SBUS<n>:I2S:CLOCk:SLOPe
(see page 1428)
Command Syntax
:SBUS<n>:I2S:CLOCk:SLOPe <slope>
<slope> ::= {NEGative | POSitive}
The :SBUS<n>:I2S:CLOCk:SLOPe command specifies which edge of the I2S serial
clock signal clocks in data.
•NEGative Falling edge.
POSitive — Rising edge.
Query Syntax
:SBUS<n>:I2S:CLOCk:SLOPe?
The :SBUS<n>:I2S:CLOCk:SLOPe? query returns the current I2S clock slope
setting.
Return Format
<slope><NL>
<slope> ::= {NEG | POS}
See Also "Introduction to :TRIGger Commands" on page 1139
":SBUS<n>:I2S:ALIGnment" on page 845
":SBUS<n>:I2S:RWIDth" on page 848
":SBUS<n>:I2S:TWIDth" on page 860