:SAVE Commands 28
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 747
:SAVE:WAVeform[:STARt]
(see page 1428)
Command Syntax
:SAVE:WAVeform[:STARt] [<file_name>]
<file_name> ::= quoted ASCII string
The :SAVE:WAVeform[:STARt] command saves oscilloscope waveform data to a
file.
See Also • "Introduction to :SAVE Commands" on page 725
• ":SAVE:WAVeform:FORMat" on page 748
• ":SAVE:WAVeform:LENGth" on page 749
• ":SAVE:FILename" on page 729
• ":RECall:SETup[:STARt]" on page 721
Be sure to set the :SAVE:WAVeform:FORMat before saving waveform data. If the format is
NONE, the save waveform command will not succeed.
If a file extension is provided as part of a specified <file_name>, and it does not match the
extension expected by the format specified in :SAVE:WAVeform:FORMat, the format will be
changed if the extension is a valid waveform file extension.