:SBUS<n> Commands 29
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 929
:SBUS<n>:SENT:TRIGger:TOLerance
(see page 1428)
Command Syntax
:SBUS<n>:SENT:TRIGger:TOLerance <percent>
<percent> ::= from 1-28, in NR1 format.
The :SBUS<n>:SENT:TRIGger:TOLerance command specifies the tolerance
variation that is considered a violation.
The trigger tolerance can be up to the :SBUS<n>:SENT:TOLerance setting minus
two percent. For example, after sending ":SBUS1:SENT:TOLerance 20", the
:SBUS1:SENT:TRIGger:TOLerance setting can be from 1 to 18.
Query Syntax
:SBUS<n>:SENT:TRIGger:TOLerance?
The :SBUS<n>:SENT:TRIGger:TOLerance? query returns tolerance variation
percent setting.
Return Format
<percent><NL>
<percent> ::= from 1-28, in NR1 format.
See Also ":SBUS<n>:SENT:CLOCk" on page 900
":SBUS<n>:SENT:CRC" on page 901
":SBUS<n>:SENT:DISPlay" on page 902
":SBUS<n>:SENT:FORMat" on page 904
":SBUS<n>:SENT:IDLE" on page 906
":SBUS<n>:SENT:LENGth" on page 907
":SBUS<n>:SENT:PPULse" on page 908
":SBUS<n>:SENT:SIGNal<s>:DISPlay" on page 909
":SBUS<n>:SENT:SIGNal<s>:LENGth" on page 910
":SBUS<n>:SENT:SIGNal<s>:MULTiplier" on page 912
":SBUS<n>:SENT:SIGNal<s>:OFFSet" on page 913
":SBUS<n>:SENT:SIGNal<s>:ORDer" on page 914
":SBUS<n>:SENT:SIGNal<s>:STARt" on page 916
":SBUS<n>:SENT:SOURce" on page 918
":SBUS<n>:SENT:TOLerance" on page 920
":SBUS<n>:SENT:TRIGger" on page 921
":SBUS<n>:SENT:TRIGger:FAST:DATA" on page 923
":SBUS<n>:SENT:TRIGger:SLOW:DATA" on page 924
":SBUS<n>:SENT:TRIGger:SLOW:ID" on page 926
":SBUS<n>:SENT:TRIGger:SLOW:ILENgth" on page 928