:SBUS<n> Commands 29
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 787
:SBUS<n>:CAN:FDSPoint
(see page 1428)
Command Syntax
:SBUS<n>:CAN:FDSPoint <value>
<value> ::= even numbered percentages from 30 to 90 in NR3 format.
The :SBUS<n>:CAN:FDSPoint command sets the point during the bit time where
the bit level is sampled to determine whether the bit is dominant or recessive. The
sample point represents the percentage of time between the beginning of the bit
time to the end of the bit time.
Query Syntax
:SBUS<n>:CAN:FDSPoint?
The :SBUS<n>:CAN:FDSPoint? query returns the current CAN FD sample point
setting.
Return Format
<value><NL>
<value> ::= even numbered percentages from 30 to 90 in NR3 format.
See Also ":SBUS<n>:CAN:SIGNal:FDBaudrate" on page 792