:ACQuire Commands 8
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 311
:ACQuire:SEGMented:ANALyze
(see page 1610)
Command Syntax
:ACQuire:SEGMented:ANALyze
This command calculates measurement statistics and/or infinite persistence over
all segments that have been acquired. It corresponds to the front panel Analyze
Segments softkey which appears in both the Measurement Statistics and
Segmented Memory Menus.
In order to use this command, the oscilloscope must be stopped and in segmented
acquisition mode, with either quick measurements or infinite persistence on.
See Also • ":ACQuire:MODE" on page 307
• ":ACQuire:SEGMented:COUNt" on page 312
• "Introduction to :ACQuire Commands" on page 300