EasyManua.ls Logo

Keysight Technologies InfiniiVision MSO-X 4154A

Keysight Technologies InfiniiVision MSO-X 4154A
1580 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Obsolete and Discontinued Commands 37
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1371
:MTESt:RUMode
(see page 1428)
Command Syntax
:MTESt:RUMode {FORever | TIME,<seconds> | {WAVeforms,<wfm_count>}}
<seconds> ::= from 1 to 86400 in NR3 format
<wfm_count> ::= number of waveforms in NR1 format
from 1 to 1,000,000,000
The :MTESt:RUMode command determines the termination conditions for the
mask test. The choices are FORever, TIME, or WAVeforms.
FORever — runs the Mask Test until the test is turned off.
TIME — sets the amount of time in seconds that a mask test will run before it
terminates. The <seconds> parameter is a real number from 1 to 86400
seconds.
WAVeforms — sets the maximum number of waveforms that are required before
the mask test terminates. The <wfm_count> parameter indicates the number of
waveforms that are to be acquired; it is an integer from 1 to 1,000,000,000.
Query Syntax
:MTESt:RUMode?
The :MTESt:RUMode? query returns the currently selected termination condition
and value.
Return Format
{FOR | TIME,<seconds> | {WAV,<wfm_count>}}<NL>
<seconds> ::= from 1 to 86400 in NR3 format
<wfm_count> ::= number of waveforms in NR1 format
from 1 to 1,000,000,000
See Also "Introduction to :MTESt Commands" on page 581
":MTESt:RUMode:SOFailure" on page 1372
NOTE
The :MTESt:RUMode command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MTESt:RMODe command (see page 598) instead.

Table of Contents

Related product manuals