:FUNCtion<m> Commands 18
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 415
:FUNCtion<m>:OFFSet
(see page 1428)
Command Syntax
:FUNCtion<m>:OFFSet <offset>
<m> ::= 1 to (# math functions) in NR1 format
<offset> ::= the value at center screen in NR3 format.
The :FUNCtion<m>:OFFSet command sets the voltage or vertical value
represented at center screen for the selected function. The range of legal values is
generally +/-10 times the current scale of the selected function, but will vary by
function. If you set the offset to a value outside of the legal range, the offset value
is automatically set to the nearest legal value.
Query Syntax
:FUNCtion<m>:OFFSet?
The :FUNCtion<m>:OFFSet? query outputs the current offset value for the selected
function.
Return Format
<offset><NL>
<offset> ::= the value at center screen in NR3 format.
See Also "Introduction to :FUNCtion<m> Commands" on page 393
":FUNCtion<m>:RANGe" on page 420
":FUNCtion<m>:REFerence" on page 421
":FUNCtion<m>:SCALe" on page 422
NOTE
The :FUNCtion<m>:OFFSet command is equivalent to the :FUNCtion<m>:REFerence
command.