838 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:FLEXray:TRIGger:EVENt:TYPE
(see page 1428)
Command Syntax
:SBUS<n>:FLEXray:TRIGger:EVENt:TYPE <event>
<event> ::= {WAKeup | TSS | {FES | DTS} | BSS}
Selects the FlexRay event to trigger on. The event setting is only valid when the
FlexRay trigger mode is set to EVENt.
WAKeup — triggers on Wake-Up event.
TSS — triggers on Transmission Start Sequence event.
FES — triggers on either Frame End or Dynamic Trailing Sequence event.
DTS — triggers on either Frame End or Dynamic Trailing Sequence event.
BSS — triggers on Byte Start Sequence event.
Query Syntax
:SBUS<n>:FLEXray:TRIGger:EVENt:TYPE?
The :SBUS<n>:FLEXray:TRIGger:EVENt:TYPE? query returns the currently selected
FLEXray event.
Return Format
<event><NL>
<event> ::= {WAK | TSS | {FES | DTS} | BSS}
See Also "Introduction to :TRIGger Commands" on page 1139
":SBUS<n>:FLEXray:TRIGger:EVENt:AUToset" on page 836
":SBUS<n>:FLEXray:TRIGger:EVENt:BSS:ID" on page 837
":SBUS<n>:FLEXray:TRIGger" on page 834
":SBUS<n>:FLEXray:AUTosetup" on page 826
":SBUS<n>:FLEXray:SOURce" on page 833