EasyManua.ls Logo

Keysight Technologies InfiniiVision MSO-X 4154A

Keysight Technologies InfiniiVision MSO-X 4154A
1580 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
568 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
23 :MEASure Power Commands
:MEASure:ONTime
(see page 1428)
Command Syntax
:MEASure:ONTime [<source1>][,<source2>]
<source1>, <source2> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:ONTime command installs an "on time" measurement on screen.
Turn on time measures the difference of time between when the input AC Voltage
first rises to 10% of its maximum amplitude to the time when the output DC
Voltage rises to 90% of its maximum amplitude.
The <source1> parameter is the AC Voltage and the <source2> parameter is the
DC Voltage. These sources can also be specified by the :MEASure:SOURce
command.
Query Syntax
:MEASure:ONTime? [<source1>][,<source2>]
The :MEASure:ONTime query returns the measured turn off time.
Return Format
<return_value><NL>
<return_value> ::= the time in seconds in NR3 format
See Also ":MEASure:SOURce" on page 529
":POWer:ONOFf:TEST" on page 664
":POWer:ONOFf:APPLy" on page 661

Table of Contents

Related product manuals