1298 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
35 :WGEN<w> Commands
:WGEN<w>:MODulation:STATe
(see page 1428)
Command Syntax
:WGEN<w>:MODulation:STATe <setting>
<w> ::= 1 in NR1 format
<setting> ::= {{OFF | 0} | {ON | 1}}
The :WGEN<w>:MODulation:STATe command enables or disables modulated
waveform generator output.
You can enable modulation for all waveform generator function types except
pulse, DC, and noise.
Query Syntax
:WGEN<w>:MODulation:STATe?
The :WGEN<w>:MODulation:STATe? query returns whether the modulated
waveform generator output is enabled of disabled.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also ":WGEN<w>:MODulation:AM:DEPTh" on page 1289
":WGEN<w>:MODulation:AM:FREQuency" on page 1290
":WGEN<w>:MODulation:FM:DEViation" on page 1291
":WGEN<w>:MODulation:FM:FREQuency" on page 1292
":WGEN<w>:MODulation:FSKey:FREQuency" on page 1293
":WGEN<w>:MODulation:FSKey:RATE" on page 1294
":WGEN<w>:MODulation:FUNCtion" on page 1295
":WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry" on page 1296
":WGEN<w>:MODulation:TYPE" on page 1299