:SEARch Commands 32
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1223
:SEARch:SERial:LIN:SYMBolic:VALue
(see page 1610)
Command Syntax
:SEARch:SERial:LIN:SYMBolic:VALue <data>
<data> ::= value in NR3 format
The :SEARch:SERial:LIN:SYMBolic:VALue command specifies the signal value to
search for when LIN symbolic data has been loaded (recalled) into the oscilloscope
and the LIN serial search mode is set to FSIGnal.
Query Syntax
:SEARch:SERial:LIN:SYMBolic:VALue?
The :SEARch:SERial:LIN:SYMBolic:VALue? query returns the specified signal value.
Return Format
<data><NL>
<data> ::= value in NR3 format
See Also • ":RECall:LDF[:STARt]" on page 815
• ":SEARch:SERial:LIN:MODE" on page 1217
• ":SEARch:SERial:LIN:SYMBolic:FRAMe" on page 1221
• ":SEARch:SERial:LIN:SYMBolic:SIGNal" on page 1222
NOTE
Encoded signal values are not supported in the remote interface (even though they can be
used in the front panel graphical interface).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision MSO-X 4154A and is the answer not in the manual?

Keysight Technologies InfiniiVision MSO-X 4154A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision MSO-X 4154A
CategoryTest Equipment
LanguageEnglish

Related product manuals