:POWer Commands 26
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 629
:POWer:CLResponse:FREQuency:MODE
(see page 1428)
Command Syntax
:POWer:CLResponse:FREQuency:MODE <mode>
<mode> ::= {SWEep | SINGle}
The :POWer:CLResponse:FREQuency:MODE command specifies whether the
analysis should be performed by sweeping through a range of frequencies
(SWEep) or at a single frequency (SINGle).
The SINGle mode is useful for evaluating amplitudes at a single frequency, for
example, near the expected 0 dB cross-over frequency. After running the test at a
single frequency, you can manually adjust (increase) the waveform generator's
amplitude until you begin to observe distortion in the waveforms on the
oscilloscope's display. You can then use that amplitude at all frequencies in
SWEep mode, or you can evaluate amplitudes at other frequencies in order to
determine an optimized amplitude profile (see
:POWer:CLResponse:WGEN:VOLTage:PROFile).
Query Syntax
:POWer:CLResponse:FREQuency:MODE?
The :POWer:CLResponse:FREQuency:MODE? query returns the frequency mode
setting.
Return Format
<mode><NL>
<mode> ::= {SWE | SING}
See Also • ":POWer:CLResponse" on page 626
• ":POWer:CLResponse:APPLy" on page 627
• ":POWer:CLResponse:DATA" on page 628
• ":POWer:CLResponse:FREQuency:STARt" on page 630
• ":POWer:CLResponse:FREQuency:STOP" on page 631
• ":POWer:CLResponse:PPDecade" on page 632
• ":POWer:CLResponse:SOURce:INPut" on page 633
• ":POWer:CLResponse:SOURce:OUTPut" on page 634
• ":POWer:CLResponse:WGEN:LOAD" on page 635
• ":POWer:CLResponse:WGEN:VOLTage" on page 636
• ":POWer:CLResponse:WGEN:VOLTage:PROFile" on page 637