636 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
26 :POWer Commands
:POWer:CLResponse:WGEN:VOLTage
(see page 1428)
Command Syntax
:POWer:CLResponse:WGEN:VOLTage <amplitude>[,<range>]
<amplitude> ::= amplitude in volts in NR3 format
<range> ::= {F20HZ | F100HZ | F1KHZ | F10KHZ | F100KHZ | F1MHZ
| F10MHZ | F20MHZ}
The :POWer:CLResponse:WGEN:VOLTage command sets the waveform generator
output amplitude(s).
When the waveform generator amplitude profile is enabled (with the
:POWer:CLResponse:WGEN:VOLTage:PROFile command), you can set an initial
ramp amplitude for each frequency range.
Without the <range> parameter, this command sets the waveform generator
output amplitude used when the amplitude profile is disabled.
Query Syntax
:POWer:PSRR:WGEN:VOLTage? [<range>]
The :POWer:CLResponse:WGEN:VOLTage? query returns the waveform generator
output amplitude setting(s).
Return Format
<amplitude><NL>
<amplitude> ::= amplitude in volts in NR3 format
See Also ":POWer:CLResponse" on page 626
":POWer:CLResponse:APPLy" on page 627
":POWer:CLResponse:DATA" on page 628
":POWer:CLResponse:FREQuency:MODE" on page 629
":POWer:CLResponse:FREQuency:STARt" on page 630
":POWer:CLResponse:FREQuency:STOP" on page 631
":POWer:CLResponse:PPDecade" on page 632
":POWer:CLResponse:SOURce:INPut" on page 633
":POWer:CLResponse:SOURce:OUTPut" on page 634
":POWer:CLResponse:WGEN:LOAD" on page 635
":POWer:CLResponse:WGEN:VOLTage:PROFile" on page 637