866 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:IIC:TRIGger:PATTern:ADDRess
(see page 1428)
Command Syntax
:SBUS<n>:IIC:TRIGger:PATTern:ADDRess <value>
<value> ::= integer or <string>
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F}
The :SBUS<n>:IIC:TRIGger:PATTern:ADDRess command sets the address for IIC
data.The address can range from 0x00 to 0x7F (7-bit) or 0x3FF (10-bit)
hexadecimal. Use the don't care address (-1 or 0xFFFFFFFF) to ignore the address
value.
Query Syntax
:SBUS<n>:IIC:TRIGger:PATTern:ADDRess?
The :SBUS<n>:IIC:TRIGger:PATTern:ADDRess? query returns the current address
for IIC data.
Return Format
<value><NL>
<value> ::= integer
See Also "Introduction to :TRIGger Commands" on page 1139
":SBUS<n>:IIC:TRIGger:PATTern:DATA" on page 867
":SBUS<n>:IIC:TRIGger:PATTern:DATa2" on page 868
":SBUS<n>:IIC:TRIGger[:TYPE]" on page 870