EasyManua.ls Logo

Keysight Technologies InfiniiVision MSO-X 4154A

Keysight Technologies InfiniiVision MSO-X 4154A
1580 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:TRIGger Commands 33
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1183
:TRIGger:NFC:STANdard
(see page 1428)
Command Syntax
:TRIGger:NFC:STANdard <standard>
<standard> ::= {{A | A106} | {B | B106} | F212 | F424}
The :TRIGger:NFC:STANdard command selects the signaling technology used by
the input signal:
A or A106 — NFC-A standard, 106 kbits/s.
B or B106 — NFC-A standard, 106 kbits/s.
F212 — NFC-F standard, 212 kbits/s.
F424 — NFC-F standard, 424 kbits/s.
Query Syntax
:TRIGger:NFC:STANdard?
The :TRIGger:NFC:STANdard? query returns the signaling technology setting.
Return Format
<standard><NL>
<standard> ::= {{A | A106} | {B | B106} | F212 | F424}
See Also ":TRIGger:NFC:AEVent" on page 1180
":TRIGger:NFC:ATTime" on page 1181
":TRIGger:NFC:SOURce" on page 1182
":TRIGger:NFC:TEVent" on page 1184
":TRIGger:NFC:TIMeout" on page 1186
":TRIGger:NFC:TIMeout:ENABle" on page 1187
":TRIGger:NFC:TIMeout:TIME" on page 1188

Table of Contents