:MTESt Commands 24
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 585
:MTESt:AMASk:CREate
(see page 1428)
Command Syntax
:MTESt:AMASk:CREate
The :MTESt:AMASk:CREate command automatically constructs a mask around the
current selected channel, using the tolerance parameters defined by the
:MTESt:AMASk:XDELta, :MTESt:AMASk:YDELta, and :MTESt:AMASk:UNITs
commands. The mask only encompasses the portion of the waveform visible on
the display, so you must ensure that the waveform is acquired and displayed
consistently to obtain repeatable results.
The :MTESt:SOURce command selects the channel and should be set before using
this command.
See Also "Introduction to :MTESt Commands" on page 581
":MTESt:AMASk:XDELta" on page 588
":MTESt:AMASk:YDELta" on page 589
":MTESt:AMASk:UNITs" on page 587
":MTESt:AMASk:SOURce" on page 586
":MTESt:SOURce" on page 611
Example Code "Example Code" on page 581