:TIMebase Commands 32
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1129
:TIMebase:RANGe
(see page 1428)
Command Syntax
:TIMebase:RANGe <range_value>
<range_value> ::= time for 10 div in seconds in NR3 format
The :TIMebase:RANGe command sets the full-scale horizontal time in seconds for
the main window. The range is 10 times the current time-per-division setting.
Query Syntax
:TIMebase:RANGe?
The :TIMebase:RANGe query returns the current full-scale range value for the
main window.
Return Format
<range_value><NL>
<range_value> ::= time for 10 div in seconds in NR3 format
See Also "Introduction to :TIMebase Commands" on page 1126
":TIMebase:MODE" on page 1127
":TIMebase:SCALe" on page 1133
":TIMebase:WINDow:RANGe" on page 1136
Example Code
' TIME_RANGE - Sets the full scale horizontal time in seconds. The
' range value is 10 times the time per division.
myScope.WriteString ":TIM:RANG 2e-3" ' Set the time range to 0.002
seconds.
See complete example programs at: Chapter 42, “Programming Examples,”
starting on page 1437