:SBUS<n> Commands 29
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 819
:SBUS<n>:CXPI:TRIGger:PATTern:DATA:STARt
(see page 1428)
Command Syntax
:SBUS<n>:CXPI:TRIGger:PATTern:DATA:STARt <start>
<start> ::= integer between 0 and 255, in NR1 format.
When triggering on long frames (with the LDATa trigger type) that can have up to
255 data bytes, the maximum number of data bytes you can include in the trigger
specification is still only 12 bytes. In this case, you can use the
:SBUS<n>:CXPI:TRIGger:PATTern:DATA:STARt command to specify the starting
byte location where the data value should be found.
The :SBUS<n>:CXPI:TRIGger:PATTern:DATA:LENGth command lets you specify the
length of the data value to trigger on.
The :SBUS<n>:CXPI:TRIGger:PATTern:DATA command lets you specify the data
value to trigger on.
Query Syntax
:SBUS<n>:CXPI:TRIGger:PATTern:DATA:STARt?
The :SBUS<n>:CXPI:TRIGger:PATTern:DATA:STARt? query returns the start byte
setting.
Return Format
<start><NL>
<start> ::= integer between 0 and 255, in NR1 format.
See Also ":SBUS<n>:CXPI:TRIGger" on page 813
":SBUS<n>:CXPI:TRIGger:IDFilter" on page 815
":SBUS<n>:CXPI:TRIGger:PTYPe" on page 816
":SBUS<n>:CXPI:TRIGger:PATTern:DATA" on page 817
":SBUS<n>:CXPI:TRIGger:PATTern:DATA:LENGth" on page 818
":SBUS<n>:CXPI:TRIGger:PATTern:ID" on page 820
":SBUS<n>:CXPI:TRIGger:PATTern:INFO:CT" on page 821
":SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC" on page 822
":SBUS<n>:CXPI:TRIGger:PATTern:INFO:NM" on page 823