694 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
26 :POWer Commands
:POWer:SIGNals:VMAXimum:ONOFf:ON
(see page 1428)
Command Syntax
:POWer:SIGNals:VMAXimum:ONOFf:ON <value>[suffix]
<value> ::= Maximum expected input Voltage in NR3 format
[suffix] ::= {V | mV}
The :POWer:SIGNals:VMAXimum:ONOFf:ON command specifies the maximum
expected input voltage. This value is used to set the vertical scale of the channel
probing voltage for turn on analysis.
When the :POWer:ITYPe is DC, this command defines the maximum DC input
voltage amplitude value. The values can be negative.
When the :POWer:ITYPe is AC, this command defines the maximum peak-to-peak
input voltage. Only positive values are allowed.
Query Syntax
:POWer:SIGNals:VMAXimum:ONOFf:ON?
The :POWer:SIGNals:VMAXimum:ONOFf:ON query returns the expected maximum
input voltage setting.
Return Format
<value><NL>
<value> ::= Maximum expected input Voltage in NR3 format
See Also ":POWer:ITYPe" on page 657
":POWer:ONOFf:APPLy" on page 661
":POWer:SIGNals:AUTosetup" on page 681
":POWer:SIGNals:DURation:ONOFf:ON" on page 687
":POWer:SIGNals:IEXPected" on page 690
":POWer:SIGNals:OVERshoot" on page 691
":POWer:SIGNals:VSTeady:ONOFf:ON" on page 696
":POWer:SIGNals:SOURce:CURRent<i>" on page 698
":POWer:SIGNals:SOURce:VOLTage<i>" on page 699