602 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
24 :MTESt Commands
:MTESt:RMODe:FACTion:STOP
(see page 1428)
Command Syntax
:MTESt:RMODe:FACTion:STOP <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:STOP command sets stopping on a mask failure on
or off. When this setting is ON and a mask violation is detected, the mask test is
stopped and the acquisition system is stopped.
Query Syntax
:MTESt:RMODe:FACTion:STOP?
The :MTESt:RMODe:FACTion:STOP? query returns the current mask failure stop
setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 581
":MTESt:RMODe:FACTion:MEASure" on page 599
":MTESt:RMODe:FACTion:PRINt" on page 600