570 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
23 :MEASure Power Commands
:MEASure:PCURrent
(see page 1428)
Command Syntax
:MEASure:PCURrent [<source>]
<source> ::= {CHANnel<n>| FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:PCURrent command installs a peak current measurement on
screen.
The <source> parameter is the channel probing the current. This source can also
be specified by the :MEASure:SOURce command.
This command measures the peak current when the power supply first turned on.
Query Syntax
:MEASure:PCURrent? [<source>]
The :MEASure:PCURrent query returns the measured peak current.
Return Format
<return_value><NL>
<return_value> ::= the peak current value in NR3 format
See Also ":MEASure:SOURce" on page 529
":POWer:INRush:APPLy" on page 654