EasyManua.ls Logo

Keysight Technologies InfiniiVision MSO-X 4154A

Keysight Technologies InfiniiVision MSO-X 4154A
1580 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1144 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
33 :TRIGger Commands
:TRIGger:HOLDoff
(see page 1428)
Command Syntax
:TRIGger:HOLDoff <holdoff_time>
<holdoff_time> ::= 40 ns to 10 s in NR3 format
The :TRIGger:HOLDoff command defines the holdoff time value in seconds.
Holdoff keeps a trigger from occurring until after a certain amount of time has
passed since the last trigger. This feature is valuable when a waveform crosses the
trigger level multiple times during one period of the waveform. Without holdoff,
the oscilloscope could trigger on each of the crossings, producing a confusing
waveform. With holdoff set correctly, the oscilloscope always triggers on the same
crossing. The correct holdoff setting is typically slightly less than one period.
Query Syntax
:TRIGger:HOLDoff?
The :TRIGger:HOLDoff? query returns the holdoff time value for the current trigger
mode.
Return Format
<holdoff_time><NL>
<holdoff_time> ::= the holdoff time value in seconds in NR3 format.
See Also "Introduction to :TRIGger Commands" on page 1139