1460 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
37 :WGEN<w> Commands
:WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry
(see page 1610)
Command Syntax
:WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry <percent>
<w> ::= 1 in NR1 format
<percent> ::= symmetry percentage from 0% to 100% in NR1 format
The :WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry command specifies the
amount of time per cycle that the ramp waveform is rising. The ramp modulating
waveform shape is specified with the :WGEN<w>:MODulation:FUNCtion
command.
Query Syntax
:WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry?
The :WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry? query returns ramp
symmetry percentage setting.
Return Format
<percent><NL>
<percent> ::= symmetry percentage from 0% to 100% in NR1 format
See Also • ":WGEN<w>:MODulation:AM:DEPTh" on page 1453
• ":WGEN<w>:MODulation:AM:FREQuency" on page 1454
• ":WGEN<w>:MODulation:FM:DEViation" on page 1455
• ":WGEN<w>:MODulation:FM:FREQuency" on page 1456
• ":WGEN<w>:MODulation:FSKey:FREQuency" on page 1457
• ":WGEN<w>:MODulation:FSKey:RATE" on page 1458
• ":WGEN<w>:MODulation:FUNCtion" on page 1459
• ":WGEN<w>:MODulation:STATe" on page 1462
• ":WGEN<w>:MODulation:TYPE" on page 1463

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision MSO-X 4154A and is the answer not in the manual?

Keysight Technologies InfiniiVision MSO-X 4154A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision MSO-X 4154A
CategoryTest Equipment
LanguageEnglish

Related product manuals