922 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC
(see page 1610)
Command Syntax
:SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC <dlc>
<dlc> ::= integer between -1 (don't care) and 15, in NR1 format,
when trigger is in DATA mode.
<dlc> ::= integer between -1 (don't care) and 255, in NR1 format,
when trigger is in LDATa mode.
For the trigger types that let you trigger on data, the
:SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC command specifies the data length
code of the CXPI frame you wish to trigger on.
This will also affect the number of data bytes you can specify in the trigger.
Query Syntax
:SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC?
The :SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC? query returns the DLC trigger
value setting.
Return Format
<dlc><NL>
<dlc> ::= integer between -1 (don't care) and 15, in NR1 format,
when trigger is in DATA mode.
<dlc> ::= integer between -1 (don't care) and 255, in NR1 format,
when trigger is in LDATa mode.
See Also • ":SBUS<n>:CXPI:TRIGger" on page 913
• ":SBUS<n>:CXPI:TRIGger:IDFilter" on page 915
• ":SBUS<n>:CXPI:TRIGger:PTYPe" on page 916
• ":SBUS<n>:CXPI:TRIGger:PATTern:DATA" on page 917
• ":SBUS<n>:CXPI:TRIGger:PATTern:DATA:LENGth" on page 918
• ":SBUS<n>:CXPI:TRIGger:PATTern:DATA:STARt" on page 919
• ":SBUS<n>:CXPI:TRIGger:PATTern:ID" on page 920
• ":SBUS<n>:CXPI:TRIGger:PATTern:INFO:CT" on page 921
• ":SBUS<n>:CXPI:TRIGger:PATTern:INFO:NM" on page 923