CALCulate Subsystem (SPECM, K14) R&S FSL
1300.2519.12 6.396 E-11
CALCulate:MARKer:SPECtrogram Subsystem (SPECM, K14)
The CALCulate:MARKer:SPECtrogram subsystem controls the marker functions of the Spectrogram
Measurement option.
Commands of the CALCulate:MARKer:SPECtrogram Subsystem
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:FRAMe
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:SARea
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:XY:MAXimum[:PEAK]
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:XY:MINimum[:PEAK]
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:Y:MAXimum[:PEAK]
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:Y:MAXimum:ABOVe
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:Y:MAXimum:BELow
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:Y:MAXimum:NEXT
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:Y:MINimum[:PEAK]
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:Y:MINimum:ABOVe
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:Y:MINimum:BELow
– CALCulate<1|2>:MARKer<1...4>:SPECtrogram:Y:MINimum:NEXT
CALCulate<1|2>:MARKer<1...4>:SPECtrogram:FRAMe
This command defines the frame number of the specified marker.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
0 to <number of recorded frames>
If the maximum number of frames is recorded, the max. value depends on the history depth (for
details refer to CALCulate<1|2>:SPECtrogram:HDEPth command)
Example
CALC:MARK3:SPEC:FRAM –20
Positions marker 3 on frame number –20.
Characteristics
RST value: 0
SCPI: device–specific
Mode
SPECM