Machinery Balancing Module
How to Perform a 1-2 Plane with Prognosis Procedure
Perform a Correction Weight Run
Start the machine.
From the Correction Weight screen, press an Enter / Fire button to begin the
correction run. The Correction Run screen displays, showing the amount of
residual unbalance for both planes with the correction weight attached to plane A.
Figure 5 - 38.
Example of a 2-Plane Correction Run Screen When You
Select to Proceed with a 1-Plane Prognosis.
Verify whether the amount of residual unbalance is within specifications for the
machine.
If the amount of residual imbalance is within specifications, you can end the balance job,
if not, you should proceed with a trim run.
To end the balance job:
From the Correction Run screen, after you have Stopped data collection, press the
shift (zero / up arrow) Save function button. The Balance Save Setup screen
displays. Refer to Chapter 4, The Analyzer Module’s section on How to Save,
View, and Delete Analyzer Measurement Results for details on how to save
balancing jobs.
To proceed with a trim run(s) (also called check run(s)):
From the Correction Run screen, after you have Stopped data collection, press the
Start function button (or an Enter / Fire button) to proceed to the second
correction weight screen (CW 2).
Repeat the correction weight / run procedures as described above.