5. Operation sequence
57010
4/8
57010
Replenishment operation sequence [F]
10. The replenishment solution level sensor (lower) detects the replenishment solution.
• P1R (Lower) Replenishment Solution Level Sensor: OFF→ON
• P2RA (Lower) Replenishment Solution Level Sensor: OFF→ON
• P2RB (Lower) Replenishment Solution Level Sensor: OFF→ON
If any of replenishment solution level sensors (lower) does not turn on although 40 seconds has passed
after the replenishment cartridge opened
• ☞No. 05908[F] Failed to open the replenisher cartridge.
If no replenishment solution level sensors (lower) turn on although 40 seconds has passed after the
replenishment cartridge opened
• ☞No. 05905[F] Install the new replenisher cartridge.
11. The system operates the replenishment cartridge cleaning pump.
• Replenishment cartridge cleaning pump: OFF → ON
P17
P2RAP1R
PSR
R
P
2R
P
3R
P
4
P1 P2 PS-3PS-2PS-1 PS-4
P2RB
R
P
1
SE63
SE64
SE65
SE66
SE67
SE68
LSE7
LSE8
Replenisher tank
Replenishment
cartridge
G085082