3-80 Return to Section Topics 4200-900-01 Rev. K / February 2017
Section 3: Common Device Characterization Tests Model 4200-SCS User’s Manual
Figure 3-64
AC Pulse stress-measure—hardware connections
Table 3-26
Key pulse generator parameters—AC stress for WLR
Parameters Range / Specification
Rise / Fall time Variable 100 ns–500μs
Pulse width 200 ns–1ms single pulse
Pulse amplitude -5 to +5 V
Duty cycle ~50%
Base voltage +/- 5 V
Model 4200-SCS
Instrument Slots
4205
PG2
4200
SCP2HR
INSTRUMENTS
SLOT
8
SLOT
7
SLOT
6
SLOT
5
SLOT
4
SLOT
3
SLOT
2
SLOT
1
4200
SMU
SENSE LO
SENSE
FORCE
PA CNTRL
KEITHLEY
4200
SMU
SENSE LO
SENSE
FORCE
PA CNTRL
KEITHLEY
4210
SMU
SENSE LO
SENSE
FORCE
PA CNTRL
KEITHLEY
4210
SMU
SENSE LO
SENSE
FORCE
PA CNTRL
KEITHLEY
KEITHLEY
Channel 1
Channel 2
Ext
Trg
Ext
Clk
Row E
A = Adapter, BNC Male to 3-lug Triax Male
B = Adapter, SMA Female to BNC Female
Model 7072 (side view to show
connector adapters)
NOTE Use torque wrench to tighten SMA
connections to
8 inch-lbs.
Scope Card
Model 4205-PG2
Model 4200-SMU (1)
Model 4200-SMU (2)
A
To Model 4205-PG2 Channel 1
708A SWITCHING SYSTEM
KEITHLEY
A
B
C
D
E
F
G
H
1 2 3 4 5 6 7 8 9 10 11 12
TALK
LSTN
REM
ERR
COPY
LIGHT PEN
WARNING: NO INTERNAL OPERATOR SERVICEABLE PARTS. SERVICE BY QUALIFIED SERVICE PERSONNEL ONLY
IN OUT
RELAY
DISPLAY RELAYS
DIG I/O
POWER
COPY
OPEN
RESET
LOCAL
ABCD EFGH
B
11223 45 678 91011Columns
Model 7072
Model 708A
Rows
Model 4200-SMU (3)
Drain
Gate
Substrate
Source
CHANNEL 1
KEITHLEY
TRIGGER
OUT
CHANNEL 2
TRIGGER
IN
4200-MTRX-X
Cables (2m, 6ft) (3)
White SMA Cable (2m, 6ft)
(male-to-male)
Black BNC Cable
(male-to-male)
(1 of 4)