EasyManuals Logo

Keithley 4200-SCS User Manual

Keithley 4200-SCS
335 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
4200-900-01 Rev. K / February 2017 Return to Section Topics 1-35
Model 4200-SCS User’s Manual Section 1: Getting Started
SENSE terminal
The SENSE terminal is a standard triaxial connector used to apply the ground unit SENSE signal
to the DUT in a remote sense application.
The center pin is SENSE
The inner shield is GUARD
The outer shield is circuit COMMON
When the ground unit SENSE signal is connected to a DUT, all SMU/preamp measurements will
be made relative to this DUT connection.
COMMON terminal
The COMMON terminal is a binding post that provides access to circuit COMMON.
NOTE Normally, a link is connected between ground unit COMMON and chassis ground,
but it may be necessary to remove the link to avoid measurement problems caused
by ground loops or electrical interference. Refer to the Reference Manual,
Interference, page 5-25 for details.
Chassis ground
This binding post provides a convenient connecting point to system chassis ground for the
purpose of shielding a test fixture.
Figure 1-30
Chassis ground
INSTRUMENT
CONNECTIONS
SMU ONLY
SMU AND GNDU
GNDU
SLOT
8
S
E
N
S
E
F
O
R
C
E
C
O
M
M
O
N
SENSE LO
GUARD
SENSE LO
COMMON
COMMON
FORCE
SENSE
GUARD
Chassis (Earth) Ground

Table of Contents

Other manuals for Keithley 4200-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200-SCS and is the answer not in the manual?

Keithley 4200-SCS Specifications

General IconGeneral
BrandKeithley
Model4200-SCS
CategorySemiconductors
LanguageEnglish