5-4 Return to Section Topics 4200-900-01 Rev. K / February 2017
Section 5: How to Generate Basic Pulses Model 4200-SCS User’s Manual
Standard pulse waveforms
Standard pulse waveforms are configured and controlled from the pulse card tabs in the GUI.
Figure 5-3 explains how to use the GUI for Standard Pulse output.
Standard pulse waveform previewers: KPulse provides a preview of configured standard pulse
waveforms for each enabled channel. Each waveform previewer shows the high and low levels,
and timing for the waveform.
In Figure 5-3, the configuration shown in the waveform previewer for Channel 1 uses the default
settings for KPulse (pulse high = 1 V and pulse low = 0 V). Channel 2 uses the same settings, but
the complement mode is enabled. Pulse high goes to pulse low level (0 V) and pulse low goes to
pulse high level (1 V).