Rev.1
Page 1-
31
To define the CUSTOM program to associate to one of the pre-set position (1, 2, 3 or 4) proceed as follows:
1. Select PROGRAMS, select the program number to which the program is to be associated (see the previous description) and then select CUSTOM in the next
screen; the following menu appears:
134°C PROCESS +
121°C PROCESS
EXIT
-
Select 121 °C to perform a custom program with a sterilization process at 121 °C or 134 °C for one at 134 °C. Move using the + and - keys
and confirm with the key.
2. You will then go the screen:
TIME: XX min
+/- to set
to enter
to exit
Use the + and - keys to set the duration of the sterilization process and confirm with the key.
e The duration of the sterilization process is variable from 4 to 30 minutes for the program at 134 °C, and from 20 to 30 minutes for the
program at 121 °C.
3. After selecting the time, you go to the menu where you specify the type of initial vacuum:
FRACTION. VACUUM +
SINGLE VACUUM
EXIT
-