Setup
3.4 Inputs and outputs
FST030
130 Function Manual, 04/2017, A5E35693328-AA
Bit 4: Maintenance required (MR)
Bit 5: Function check (FC)
Setting Maintenance alarm; Function check; Process value alarm; Process value warning; Mainte-
nance demanded; Maintenance required
Default
None
Related
Path
Setup → Inputs and outputs → Channel 2 - output → Status output → Alarm class
3.4.4.7.16 NAMUR status signal
Description
Selection of NAMUR status signals which impact the status output. Each diagnostic event is
allocated to a NAMUR status signal. Application if Status mode = Alarm class and Status
icons = NAMUR
Setting
Failure; Function check; Out of specification; Maintenance required
Default
None
Related
Path
Setup → Inputs and outputs → Channel 2 - output → Status output → NAMUR status signal
3.4.4.7.17 Polarity
Description
Logical polarity of status output.
Setting
Active high level; Active low level
Default
Active high level
Related
Path
Setup → Inputs and outputs → Channel 2 - output → Status output → Polarity
3.4.4.7.18 On delay
Description
Sets the time delay between alarm condition and activating the output.
Setting
0.0 s to 100.0 s
Default
0.0 s