Setup
3.2 Process values
FST030
Function Manual, 04/2017, A5E35693328-AA
55
Related
Path
Setup → Process values → Flow velocity → Custom conversion factor
3.2.4.4
Decimal places
Description
Sets the decimal places for the process value displayed in the local display.
Setting Auto; No decimal place; 1 decimal place; 2 decimal places; 3 decimal places; 4 decimal
places; 5 decimal places; 6 decimal places
Default
2 decimal places
Related
Path
Setup → Process values → Flow velocity → Decimal places
3.2.4.5
Low flow cut-off
Description
Flow velocity limit for low flow cut off. Below limit flow velocity output is forced to zero.
Setting
Min 0.00
Default
0.0 m/s
Related
Path
Setup → Process values → Flow velocity → Low flow cut-off
3.2.4.6
Alarm and warning limits
3.2.4.6.1 Upper alarm limit
Description
If process value exceeds this limit an alarm is triggered.
Setting
-9999997952 m/s to +9999997952 m/s
Default
9999997952 m/s
Related
Path
Setup → Process values Flow velocity → Alarm and warning limits → Upper alarm limit