Setup
3.1 Sensor settings
FST030
36 Function Manual, 04/2017, A5E35693328-AA
3.1.6.3
Path 2
3.1.6.3.1 Sensor frequency
Description
Desired sensor frequency for path 2. Read only.
Setting
80000.0 Hz to 4000000.0 Hz
Default
Related
Path
Setup → Sensor settings → Path settings → Path 2 → Sensor frequency
3.1.6.3.2 Path geometry
Description
The number of crossings the signal makes through the pipe (Reflect or direct geometry for
path 2).
E.g.:
Z path = 1
V path = 2
W path = 4
Setting
1 to 10
Default
1
Related
Path
Setup → Sensor settings → Path settings → Path 2 → Path geometry
3.1.6.3.3 Sensor spacing
Description
Sensor spacing - Distance between the front faces of the sensors for path 2.
Setting
-100.0 m to 100.0 m
Default
0.0 m
Related
Path
Setup → Sensor settings → Path settings → Path 2 → Sensor spacing