Setup
3.4 Inputs and outputs
FST030
164 Function Manual, 04/2017, A5E35693328-AA
Setup → Inputs and outputs → Channel 3 - relay → Input/output alarms (2)
Simulation alarms (1)
Description
Selection of alarm items that impact the output. Mainly simulation alarms. Applicable if
Status mode = Alarm item.
Setting
160 Mass flow simulated; 161 Volume flow simulated; 162 Density simulated; 163 Medium
temp. simulated; 166 Std. volume flow simulated; 167 Totalizer 1 simulated; 168 Totalizer
2 simulated; 169 Totalizer 3 simulated; 170 Loop current simulated
Setup → Inputs and outputs → Channel 3 - relay → Simulation alarms (1)
Simulation alarms (2)
Description
Selection of alarm items that impact the output. Applicable if Status mode = Alarm item.
Setting
289 Sound velocity simulated; 290 Flow velocity simulated; 291 Pressure simulated; 292
Kin. viscosity simulated; 293 Aux. temperature simulated; 294 Sensor temperature simu-
lated; 296 CH5 simulated; 297 CH6 simulated; 288 Std. volume flow simulated; 299
Standardizing factor simulated; 300 Std. Kin. viscosity simulated; 301 Std. density simulat-
ed; 302 LiquIdent simulated; 303 API gravity simulated; 304 Std. API gravity simulated;
305 Specific gravity simulated; 306 Std. spec. grav. simulated; 307 Rate of change simu-
Setup → Inputs and outputs → Channel 3 - relay → Simulation alarms (2)
Selection of alarm classes which impact the status output. Each diagnostic event is allocated
to an alarm class. Applicable if Status mode = Alarm class and Status icons = Standard.
Bit 0: Process value alarm (PA)
Bit 1: Process value warning (PW)
Bit 2: Maintenance alarm (MA)
Bit 3: Maintenance warning (MW)