Maintenance and diagnostics
4.8 Simulation
FST030
Function Manual, 04/2017, A5E35693328-AA
287
4.8.1.2.3 Simulation value
Description
Set the current value to be output on the channel when current output simulation is enabled.
Set the frequency value to be output on the channel when frequency output simulation is
enabled.
Set the pulse value to be output on the channel when pulse output simulation is enabled.
Set the signal level to be output on the channel when status output simulation is enabled.
Setting
0 mA to 25 mA
Default
0 mA
Operation mode is set to Current output
Setting 0.0 Hz to 12500.0 Hz
Default
1.0 Hz
Operation mode is set to Frequency output
Setting
0.0 pulses/s to 12500.0 pulses/s
Default
Operation mode is set to Pulse output
Setting
0; 1
Default
Operation mode is set to Status output
Related
Only if parameter
Simulation
is set to Enabled, the setting of this parameter
will be relevant.
Path
Maintenance and diagnostics → Simulation → Simulation inputs and outputs → Channel 2 -
output → Simulation value
4.8.1.3
Channel 3 - input/output
4.8.1.3.1 Operation mode
Description
Set the output functionality you want the channel to operate.